Yokogawa PC-Based MX100 User Manual

Page 70

Advertising
background image

2-41

SM MX100-01E

Testing

1

2

3

4

5

6

7

Connection

Wire the units in the same manner as in the “Output at Power ON Check.”

Settings
<MX100>

Set one channel at a time, setting the output of channels not under test to 0 V.
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.

Item

Setting

Description/Notes

Enable/disable

Enabled

Action

User output

Voltage/current

Voltage

Not relevant when using the API

Span

Need not specify

-10–10 V is acceptable.

Output upon power ON

Preset value

Sets the power ON operation as the preset.

Output upon error

Preset value

Sets the output upon error as the preset.

Preset value

+0 V

Sets +0 V as the preset value.

Other

Not required

The relationship between output values of voltage and current on the same channel are
as follows.

• 1 V 4 mA
• 5 V 20 mA

Therefore, you can check the output accuracy of voltage and current simultaneously with
one setup.

<MW100>

Referring to the MW100 user’s manual for browser operation, or by using communication
commands, , set the output at power ON according to the table below. Power cycle the
MW100 main unit or equivalent, then measure the output on all channels.
For voltage of output range

Item

Setting

Description/Notes

Output Type

AO

Output method

COM

Command output

Output range

10 V

Span

Need not specify

-10 to 10 V

Preset value

0 V

Reference channel

Not required

For current of output range

Item

Setting

Description/Notes

Output Type

AO

Output method

COM

Command output

Output range

20 mA

Span

Need not specify

0 to 20 mA

Preset value

+10 mA

Reference channel

Not required

Output Operation Setting

Item

Setting

Power ON

PRESET

Error occurence

PRESET