7 between-channel error test – Yokogawa PC-Based MX100 User Manual

Page 62

Advertising
background image

2-33

SM MX100-01E

Testing

1

2

3

4

5

6

7

Test Range and Product Specification

100 ms Interval

Test Range

Test Points

Specification

Measured Value Setting

200000 µSTR

+200000 µSTR

No. 6

±1% of Range

0 µSTR

No. 1

-200000 µSTR

No. 7

200 ms Interval

Test Range

Test Points

Specification

Measured value Setting

2000 µSTR

+2000 µSTR

No. 8

±0.5% of Range

0 µSTR

No. 1

-2000 µSTR

No. 9

20000 µSTR

+20000 µSTR

No.10

±0.3% of Range

+10000 µSTR

No,12

0 µSTR

No. 1

-10000 µSTR

No.13

-20000 µSTR

No.11

200000 µSTR

+200000 µSTR

No. 6

±0.3% of Range

0 µSTR

No. 1

-200000 µSTR

No. 7

For instructions on setting test points, see “Cal Box Settings” on the previous page.

2.7.7

Between-Channel Error Test

1.

Connect the Cal box to all channels in parallel.

2.

Restore the initial balance values (all measuring ranges) of all channels to the

initial values from a PC via communication.

3.

Enter the following settings.

Measurement range

2000 µSTR

Measurement Interval

200 ms

Filters

OFF

Duration

10 s or more

Dip switch setting (for the -B12 and -B35 modules) 4-gauge method

ON

1

2

3

4

5

SW1 to SW4: OFF
SW5: ON

4.

Apply the following input values and confirm that the between-channel error

satisfies the product specification.

Test Points

Specification

Measured value

Setting

+2000 µSTR

No. 8

+2000 µSTR ±10 µSTR

-2000 µSTR

No. 9

2000 µSTR ±10 µSTR

For instructions on setting test points, see “Cal Box Settings.”
In the case of the -B12 and -B35 modules, since the between-channel error for all
channels will be equal to the wiring resistance error, be sure to make the wiring
resistance values the same.

2.7 Tests of the 4-CH Strain Input Module