2.4.1

Insulation Resistance Test

For this test, the Four-Channel, High-Speed Universal Input Module is installed with the
MX100, MW100, or equivalent. Perform this test before the test described in section 2.4.2,
“Withstanding Voltage Test.”

procedure

Wire the instrument as shown in the figure below, measure insulation resistance, then

confirm that the results meet the criteria below.

Measured Item

Criteria

Input terminal to earth terminal

DC500 V, 10 MΩ or more

Between input terminals

DC500 V, 20 MΩ or more

Measurement between Input Terminal and Earth Terminal

Short all terminals A, B, and b (1-3) on channels 1 through 4, then measure the

insulation resistance between the MX100/MW100’s or equivalent device’s functional
earth terminals. Be sure to attach the module and base plate when performing this
test.

A 1

B 2

b 3

3ch

A 1

B 2

b 3

1ch

A 1

B 2

b 3

4ch

A 1

B 2

b 3

2ch

M

M: Insulation resistance tester

To the functional earth terminal on

the MX100, MW100, or equivalent device

Measurement between Input Terminals

Measure the insulation resistance between a device whose terminals A, B, and b (1-3)

on channels 1 and 3 are all shorted and a device whose terminals A, B, and b (1-3) on
channels 2 and 4 are all shorted.

A 1

B 2

b 3

3ch

A 1

B 2

b 3

1ch

A 1

B 2

b 3

4ch

A 1

B 2

b 3

2ch

M

M: Insuration resistance tester

2.4 Tests of the 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)