6 between-channel error test, 7 burnout test, Procedure – Yokogawa PC-Based MX100 User Manual

Page 47

Advertising
background image

2-18

SM MX100-01E

2.5.6

Between-Channel Error Test

Procedure

1.

Connect the H(A/+) side and L(B/-) side on all channels as shown in the figure below.

When connecting, do not bundle the wires on the input terminal side, but rather
near the voltage generator or resistor. Also, ensure that the resistance vales of all
three wires connected to each channel are equal. Otherwise, measurement errors
will occur.

1

2

3

1ch

1

2

3

2ch

1

2

3

3ch

1

2

3

4ch

1

2

3

1

2

3

1

2

3

1

2

3

1

2

3

1

2

3

5ch

6ch

7ch

8ch

9ch

10ch

–/B

+/A

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

+/A

–/B

b

Voltage

generator

Variable

resistor

2.

Confirm that the between-channel error vales are less than the criteria given below.

Range

Input

Criteria

20 mV

Short input H to L

0.000 mV ± 4 digits

Cu10 GE

9.036 Ω

Between-channel dispersion: 1.2°C

Measured value: 0±1.6°C (all channels)

2.5.7

Burnout Test

Procedure

1.

Connect 2 kΩ and (200 kΩ + 0.22 µF) pararel junction between H(A/+) and L(B/-)

on any two channels.

2.

Enter the following measurement ranges on the MX100/MW100.

Type-K thermocouple

Burnout ON

3.

Confirm that the measured results match the table below.

4.

Set the measurement range to “B/O OFF,” and confirm that the measured values

do not max out.

5.

Remove the 2 kΩ resistor in step 1, and connect a load (200 kΩ + 0.22 µF) in parallel.

6.

Enter the following measurement range on the MX100/MW100.

Type-K thermocouple

Burnout ON

7.

Confirm that the measured results match the list below.

8.

Set the measurement range to “B/O OFF” and confirm that the measured values

do not max out.

Connected Load

Measured Result

2 kΩ

Near Room Temperature

200 kΩ + 0.22 µF wired in parallel

Detection of an overrange or burnout

2.5 Tests of the 10-CH, Medium-Speed Universal Input Module (MX110-UNV-H10)