Ch1 ch2 ch3 ch4, Dip switch settings: 4-gauge method, Ndis terminal pin number – Yokogawa PC-Based MX100 User Manual

Page 59: Wiring diagram (ex.: ch2), 7 tests of the 4-ch strain input module

Advertising
background image

2-30

SM MX100-01E

Wiring for bridge voltage accuracy test (with -B12 and -B35)

Dip switch settings: 4-gauge method

ON

1 2 3 4 5

SW1 to SW4: OFF
SW5: ON

Wiring Diagram (Ex.: CH2)
Because the bridge voltage accuracy error will be equal the wiring resistance from the

clamp terminal to the load resistance, the effects of wiring resistance must be kept to
0.5% or less.

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

ON

1

2

3

4

5

ON

1

2

3

4

5

ON

1

2

3

4

5

ON

1

2

3

4

5

A/+V
B/ L
C/-V
D/ H

CH1

A/+V
B/ L
C/-V
D/ H

CH2

A/+V
B/ L
C/-V
D/ H

CH3

A/+V
B/ L
C/-V
D/ H

CH4

Load resistance

+

-

Waveform measuring

instrument

Wiring for bridge voltage accuracy test (with -NDI)

Ndis terminal pin number

A

G

B

C

F

E

D

Pin name

Signal name

A

B

C

D

E

F

G

P

L

N

H COM

PS NS

Wiring Diagram (Ex.: CH2)

CH1

CH2

CH3

CH4

A1002JC

Load resistance

+

-

P

N

PS
NS

Waveform measuring

instrument

2.7 Tests of the 4-CH Strain Input Module

Advertising