2 withstanding voltage test, Procedure – Yokogawa PC-Based MX100 User Manual

Page 44

Advertising
background image

2-15

SM MX100-01E

Testing

1

2

3

4

5

6

7

2.5.1

Insulation Resistance Test

Attach the MX110 10-ch Medium-Speed Universal Input Module to the MX100, MW100,
or equivalent before performing this test.

Procedure

Wire the instrument as shown in the figure below, measure insulation resistance, then
confirm that the results meet the criteria below.

Measured Item

Criteria

Input terminal to earth terminal on the MX100, MW100, or

equivalent

20 MΩ or more (at DCV 500 V)

Between input terminals

20 MΩ or more (at DCV 500 V)

Between the input terminal and

earth terminal on the MX100 or equivalent

Between input terminals

+/A 1

–/B 2

b

3

1ch

+/A 1

–/B 2

b

3

2ch

+/A 1

–/B 2

b

3

3ch

+/A 1

–/B 2

b

3

4ch

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

5ch

6ch

7ch

8ch

9ch

10ch

Insulation resistance

tester or

Withstanding voltage

tester

+/A 1

–/B 2

b

3

1ch

+/A 1

–/B 2

b

3

2ch

+/A 1

–/B 2

b

3

3ch

+/A 1

–/B 2

b

3

4ch

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

+/A 1

–/B 2

b

3

5ch

6ch

7ch

8ch

9ch

10ch

Insulation resistance

tester or

Withstanding voltage

tester

2.5.2

Withstanding Voltage Test

Attach the 10-ch Medium-Speed Universal Input Module to the MX100, MW100, or
equivalent before performing this test.

Procedure

Wire the instrument in the same manner as in section 2.5.2, “Withstanding Voltage Test,”
then confirm that the results meet the criteria below.
With the withstanding voltage test, in order to ensure continuity from the MX100 or
equivalent’s earth to the module’s heat sink, install the B8720BB onto the base plate and
fasten with screws.

Measured Item

Criteria

Input terminal to earth terminal
on the MX100, MW100, or equivalent

Current leakage of 2 mA or less at 3700 Vrms AC for 1

minute

Between input terminals

Current leakage of 1 mA or less at 1000 Vrms AC for 1

minute

2.5 Tests of the 10-CH, Medium-Speed Universal Input Module (MX110-UNV-H10)