2-47

SM MX100-01E

Testing

1

2

3

4

5

6

7

PWM Outupt Settings
<MX100 Settings>

Connect to the PC, then enter the following settings using the MX100 standard software.
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.

Item

Setting

Remarks

Enable/disable

Enabled

Action

User output

Span

Need not specify

Output upon power ON

Preset value

Sets the output at power ON as a preset value.

Output upon error

Preset value

Sets the output upon error as a preset value.

Preset value

0.000 %

Sets 0% duty.

Resolution

Depends on the test

Pulse interval

Depends on the test

Other settings

Not required

<MW100 Settings>

Using the MW100 user’s manual or communication commands, power cycle the MW100
main unit or equivalent, then measure the output on all channels.

Item

Setting

Remarks

Output type

PWM

Output method

COM

Command output

Output range

Depends on the test

Span

Need not specify

Pulse interval

Depends on the test

Preset value

0.000 %

Sets 0% duty.

Reference channel

Not required

Output Operation Setting

Item

Setting

Power ON

PRESET

Error occurence

PRESET

Criteria (for the MX100 and MW100)

If entering the pulse interval and PWM settings according to the table below, confirm that
the output pulse interval of the specified channels satisfies the criteria.
(Set for each channel. Channels not under test output 0%).

Pulse Interval

Output Setting

Criteria

1 ms

50.000 %

1 ms ±80 ns

5 ms (1 ms ×5)

50.000 %

5 ms ±400 ns

10 ms (10 ms ×1)

50.000 %

10 ms ±800 ns

100 ms (10 ms ×10)

50.000 %

100 ms ±8 s

2.10 8-CH, Medium-Speed PWM Output Module