Yokogawa PC-Based MX100 User Manual

Page 56

Advertising
background image

2-27

SM MX100-01E

Testing

1

2

3

4

5

6

7

NDIS terminal pin and signal names

A

G

B

C

F

E

D

Pin name
Signal name

A

B

C

D

E

F

G

P

L

N

H COM PS NS

Diagram for wiring between NDIS input terminals

CH1

CH2

CH3

CH4

A1002JC

A1002JC

A1002JC

A1002JC

Pin name

Signal name

E

COM

Tester or

Withstanding

voltage

tester

• As the E terminal (COM) is common to all channels, it should be excluded when

performing the between-input terminal insulation resistance or withstanding voltage
tests. Also, the E terminals on all channels can be connected to each other.

• The cable shell, and the signal wires of channels 2 and 4 are grounded to all channels.

Dip switch settings

ON

1

2

3

4

5

SW1–SW5: ON

In order to ensure continuity from the MX100, MW100, or equivalent’s earth to the heat
sink of the module under test, you must fasten the module with screws after installing
onto the MX150 base plate.