2.7 Tests of the 4-CH Strain Input Module

Tests

2.7.1 Insulation Resistance Test
2.7.2 Withstanding Voltage Test
2.7.3 Excessive Input Test
2.7.4 ID Check
2.7.5 Bridge Voltage Accuracy Test
2.7.6 Measurement accuracy
2.7.7 Between-Channel Error Test
2.7.8 Initial Balance Error Test
2.7.9 Dip Switch Operation Test

Test Instruments

The following instruments are used for testing.

Instrument

Standard

Typical Model

MX main unit or equivalent

jig

MX100 or MW100 Main Module
MX150 Base Plate

PC

That the software required for the test

is installed

Strain Cal Box

B8720XA or equivalent

Yokogawa B8720XA

DC voltage generator

DCV ±10 V

DC power

Insulation resistance tester DC500 V

TOS9201 (Kikusui)

Withstanding voltage tester AC 2 kV–4 kV

TOS9201 (Kikusui)

Tester

A device capable of performing the

continuity tests (measured voltage of

42 V or less)

-

Waveform measuring

instrument

100 kS/s, 8-bit or more

YOKOGAWA DL1640, OR, or

equivalent

Ethernet cable

Module Settings

Enter the following settings unless otheriwse indicated.

Measurement interval

200 ms

Filters

0 (OFF)

2.7.1

Insulation Resistance Test

Perform the withstanding voltage test before performing this test.
Wire the instrument as shown in the figure below, measure insulation resistance, then
confirm that the results meet the criteria below.

Measured Item

Criteria

Input terminal to earth terminal on the MX100,

MW100, or equivalent

20 MΩ or more (at DCV 500 V)

Between input terminals

10 MΩ or more (at DCV 42 V or less)

Connector shell to the ground terminal on the

MX100 or equivalent

5 or less