2 withstanding voltage test, 3 digital output test – Yokogawa PC-Based MX100 User Manual

Page 82

Advertising
background image

2-53

SM MX100-01E

Testing

1

2

3

4

5

6

7

Output terminal to earth terminal

Ch1

Ch2

Ch3

Ch4

Ch5

Ch6

Ch7

Ch8

Ch9

Ch10

NO
C
N.C.
NO
C
N.C.
NO
C
N.C.
NO
C
N.C.
NO
C
N.C.

M

To the functional earth terminal on

the MX100, MW100 or equivalent device

M: Insulation resistance tester

Short the output terminals on all channels, and measure the insulation resistance
between the functional earth terminals on the jig of the MX100, MW100, or
equivalent.

The N.C. terminals are not connected.

2.

Confirm that the criteria below are met.

Between output terminals:

DC500 V, 20 MΩ or more

Output terminal to earth terminal:

DC500 V, 20 MΩ or more

2.11.2 Withstanding Voltage Test

For this test, the Ten-Channel, Medium-Speed Digital Output Module is installed with the
MX100, MW100, or equivalent device.
When installing the DO module onto the main unit, always fasten with screws.

Procedure

1.

Perform the withstanding voltage test between each terminal in the same manner

as test item 1 of in section 2.11.1, “Insulation Resistance Test.”

2.

Confirm that the criteria below are met.

Between output terminals:

AC2300 Vrms for 1 minute, 2 mA or less

Output terminal to earth terminal:

AC2300 Vrms for 1 minute, 2 mA or less

2.11.3 Digital Output Test

For this test, the Ten-Channel, Medium-Speed Digital Output Module is installed with the
MX100, MW100, or equivalent device.

Procedure

1.

Turn ON/OFF output on the DO module using software such as the MX100

Standard Software, and measure the conduction resistance between the NO and
C terminals on the corresponding output channel using a DMM.

2.

Perform step 1 on all channels.

3.

Confirm that the criteria below are met.

Syntax

Criteria

ON

1 Ω or less

OFF

1 MΩ or higher

2.11 Tests of the 10-CH, Medium-Speed Digital Output Module (MX125-MKC-M10)

Advertising
This manual is related to the following products: