REMOTE CONTROL
99
:DISPlay:DEMod[:WINDow]:TRACe:X
[:SCALe]:RVALue
Set
Query
Description
Sets or queries the Reference value time.
Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]: RVALue
<time>
Query Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]:
RVALue?
Parameter
<time>
<NRf>
Return parameter <NR3>
Seconds
Example
:DISP:DEM:TRAC:X:RVAL 2 ms
:DISPlay:DEMod[:WINDow]:TRACe:Y[:
SCALe]:AUTO
Set
Description
Sets the vertical display scale to auto for AM/FM
demodulation.
Syntax
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]: AUTO
<integer>
Example
:DISP:DEM:TRAC:Y:AUTO
:DISPlay:DEMod[:WINDow]:TRACe:Y
[:SCALe]:PDIVision
Set
Query
Description
Sets or queries the Y-axis scale per division.
Syntax
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
PDIVision <NRf>
Query Syntax
:DISPlay:DEMod[:WINDow]:TRACe:Y[:SCALe]:
PDIVision?
Parameter
<NRf>
AM Unit: %, FM Unit: Hz
Return parameter <NR3>
AM Unit: %, FM Unit: Hz
Example
:DISP:DEM:TRAC:Y:PDIV 2.3e+1