Sense]:semask:offset<n>:stop: relative, Sense]:semask:offset<n>:stop: relative:couple – GW Instek GSP-930 Programming Manual User Manual

Page 155

Advertising
background image

REMOTE CONTROL

153

Description

Couples the Absolute Stop amplitude to the
Absolute Start amplitude for the selected offset.

Syntax

[:SENSe]:SEMask:OFFSet<n>:STOP:ABSolute:COUPle
{OFF|ON|0|1}

Query Syntax

[:SENSe]:SEMask:OFFSet<n>:STOP:ABSolute:
COUPle?

Parameter

0

Turns coupling off.

1

Turns coupling on.

OFF

Turns coupling off.

ON

Turns coupling on.

Return parameter 0

Coupling is off.

1

Coupling is on.

Example

SEM:OFFS1:STOP:ABS:COUP 0

[:SENSe]:SEMask:OFFSet<n>:STOP:
RELative

Set

Query

Description

Sets or queries the amplitude of the stop frequency
of the Relative Mask for the selected offset.

Syntax

[:SENSe]:SEMask:OFFSet<n>:STOP:RELative <ampl>

Query Syntax

[:SENSe]:SEMask:OFFSet<n>:STOP:RELative?

Parameter/
Return parameter

<ampl>

<NR3> dBc

<n>

<NR1>offset 1~5

Example

SEM:OFFS1:STOP:REL 1.5e+1

[:SENSe]:SEMask:OFFSet<n>:STOP:
RELative:COUPle

Set

Query

Description

Couples the Relative Stop amplitude to the
Relative Start amplitude for the selected offset.

Syntax

[:SENSe]:SEMask:OFFSet<n>:STOP:RELative:COUPle
{OFF|ON|0|1}