Sense]:aset:amplitude, Sense]:aset:amplitude:auto, Sense]:aset:run – GW Instek GSP-930 Programming Manual User Manual
Page 126

GSP-930 Programming Manual
124
Parameter
<freq>
<NRf>
Return parameter <NR3>
Hz
Example
:ACPR: SPAC 2.0e+6
[:SENSe]:ASET:AMPLitude
Set
Query
Description
Sets or queries the autoset amplitude floor level.
Syntax
[:SENSe]:ASET:AMPLitude <ampl>
Query Syntax
[:SENSe]:ASET:AMPLitude?
Parameter
<ampl>
<NRf> power or voltage
Return parameter <NR3>
Example
:ASET:AMPL 8.0e+1
[:SENSe]:ASET:AMPLitude:AUTO
Set
Query
Description
Sets autoset amplitude floor level to auto or
manual or queries its state.
Syntax
[:SENSe]:ASET:AMPLitude:AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:ASET:AMPLitude:AUTO?
Parameter
0
Turn autoset amplitude floor to manual.
1
Turn autoset amplitude floor to auto.
OFF
Turn autoset amplitude floor to manual.
ON
Turn autoset amplitude floor to auto.
Return parameter 0
Autoset amplitude floor is in manual.
1
Autoset amplitude floor is in auto.
Example
:ASET:AMPL:AUTO 1
[:SENSe]:ASET:RUN
Set
Description
Activates the Autoset function.
Syntax
[:SENSe]:ASET:RUN
Example
:ASET:RUN