Sense]:emifilter:state, Sense]:emifilter:bandwidth|bwidth [:resolution – GW Instek GSP-930 Programming Manual User Manual
Page 136

GSP-930 Programming Manual
134
Parameter
0
Turn the detection mode to manual (off).
1
Turn the detection mode to auto (on).
OFF
Turn the detection mode to manual (off).
ON
Turn the detection mode to auto (on).
Return parameter 0
The detection mode is set to manual.
1
The detection mode is set to automatic.
Example
:DET:AUTO ON
[:SENSe]:EMIFilter:STATe
Set
Query
Description
Turns the EMI filter on/off or queries its state.
Syntax
[:SENSe]:EMIFilter:STATe {OFF|ON|0|1}
Query Syntax
[:SENSe]:EMIFilter:STATe?
Parameter
0
Turn the EMI filter off.
1
Turn the EMI filter on.
OFF
Turn the EMI filter off.
ON
Turn the EMI filter on.
Return parameter 0
The EMI filter is off.
1
The EMI filter is on.
Example
:EMI:STAT 0
[:SENSe]:EMIFilter:BANDwidth|BWIDth
[:RESolution]
Set
Description
Sets the EMI filter bandwidth (must be set to the
exact bandwidth).
Syntax
[:SENSe]:EMIFilter:BANDwidth|BWIDth[:RESolution]
<freq>
Parameter
<freq>
<NRf>
(Only 200Hz, 9kHz, 120kHz are valid
settings)
Example
:EMIF:BAND 2.0e+2