:CALCulate:MARKer:PEAK:DATA?

Query

Description

Returns all the top 10 peak data values in CSV
format (returns the contents of the peak table).
The <csv data> data contains 10 pairs of data from
the top 10 peaks. Each pair includes the peak
frequency and the peak amplitude. There are a
total of 10 pairs of data points (20 data entries in
total) for the <csv data> data.

Query syntax

:CALCulate:MARKer:PEAK:DATA?

Return parameter <csv data>

pk#1 freq, pk#1 amp,….. pk#10 freq,
pk#10 amp.

Example

:CALCulate:MARKer:PEAK:DATA?
>1.250e+08,-5.052e+01,1.065000000e+09,…

:CALCulate:MARKer:PEAK:EXCursion

Set

Query

Description

Sets or queries the peak excursion value.

Syntax

:CALCulate:MARKer:PEAK:EXCursion <rel ampl>

Query Syntax

:CALCulate:MARKer:PEAK:EXCursion?

Parameter

<rel ampl>

Peak excursion dB (offset from
threshold)

Return parameter <NR3>

Peak excursion in dB.

Example

:CALC:MARK:PEAK:EXC 6 db

:CALCulate:MARKer:PEAK:SORT:TYPE

Set

Query

Description

Sets or queries the peak sort type for the peak
table.

Syntax

:CALCulate:MARKer:PEAK:SORT:TYPE
{FREQuency|AMPLitude}

Query Syntax

:CALCulate:MARKer:PEAK:SORT:TYPE?