Sense]:demod:ifbw, Sense]:detector[:function, Sense]:detector[:function]:auto – GW Instek GSP-930 Programming Manual User Manual
Page 135

REMOTE CONTROL
133
Example
:DEM:FILT:LPAS B
[:SENSe]:DEMod:IFBW
Set
Query
Description
Sets or queries the IF bandwidth for the AM/FM
Analysis function.
Syntax
[:SENSe]:DEMod:IFBW <freq>
Query Syntax
[:SENSe]:DEMod:IFBW?
Parameter
<freq>
<NRf>
Return parameter <NR3>
Hz
Example
DEM:IFBW 3.0e+5
[:SENSe]:DETector[:FUNCtion]
Set
Query
Description
Sets or queries the trace detection mode when in
manual mode.
Syntax
[:SENSe]:DETector[:FUNCtion]
{AVERage|SAMPle|POSitive|NEGative|NORMal}
Query Syntax
[:SENSe]:DETector[:FUNCtion]?
Parameter/
Return parameter
AVERage
Sets the detector mode to Average.
SAMPle
Sets the detector mode to Sample.
POSitive
Sets the detector mode to Peak+.
NEGative
Sets the detector mode to Peak-.
NORMal
Sets the detector mode to Normal.
Example
:DET NORM
[:SENSe]:DETector[:FUNCtion]:AUTO
Set
Query
Description
Turns the trace detection mode to auto (on) or
manual (off) or queries its state.
Syntax
[:SENSe]:DETector[:FUNCtion]:AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:DETector[:FUNCtion]:AUTO?