Calculate:marker:peak:table:state, Calculate:marker:peak:threshold, Calculate:marker:peak:threshold: state – GW Instek GSP-930 Programming Manual User Manual
Page 86

GSP-930 Programming Manual
84
Parameter/
Return parameter
FREQuency Sort by frequency.
AMPLitude Sort by amplitude.
Example
:CALC:MARK:PEAK:SORT:TYPE FREQ
:CALCulate:MARKer:PEAK:TABLe:STATe
Set
Query
Description
Sets or queries the state of the Peak Table.
Syntax
:CALCulate:MARKer:PEAK:TABLe:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:MARKer:PEAK:TABLe:STATe?
Parameter
0
Turn peak table off.
1
Turn peak table on.
OFF
Turn peak table off.
ON
Turn peak table on.
Return parameter 0
peak table is off.
1
peak table is on.
Example
:CALC:MARK:PEAK:TABL:STAT ON
:CALCulate:MARKer:PEAK:THReshold
Set
Query
Description
Sets or queries the peak threshold value.
Syntax
:CALCulate:MARKer:PEAK:THReshold < ampl>
Query Syntax
:CALCulate:MARKer:PEAK:THReshold?
Parameter
< ampl>
Peak Threshold level
Return parameter <NR3>
Peak threshold. Note: the unit returned
depends on the currently set vertical
units.
Example
:CALC:MARK:PEAK:THR -3 dBm
:CALCulate:MARKer:PEAK:THReshold:
STATe
Set
Query
Description
Sets or queries the state of the Peak Threshold.