Configure commands, Calculate:toi:state, Configure:mode – GW Instek GSP-930 Programming Manual User Manual
Page 97

REMOTE CONTROL
95
Query syntax
:CALCulate:TOI:RESult?
Return parameter <3rd lower>
3rd order lower intercept
<lower p/f>
3
rd
order lower pass/fail judgment.
0=pass, 1=fail
<3
rd
upper>
3
rd
order upper intercept
<upper p/f>
3
rd
order lower pass/fail judgment
0=pass, 1=fail
Example
:CALC:TOI:RES?
>-5.5e+1,0, -6.61e+1,0
:CALCulate:TOI:STATe
Set
Query
Description
Turns TOI measurement on/off or queries its state.
Syntax
:CALCulate:TOI:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:TOI:STATe?
Parameter
0
Turns TOI measurement off.
1
Turns TOI measurement on.
OFF
Turns TOI measurement off.
ON
Turns TOI measurement on.
Return parameter 0
TOI measurement is off.
1
TOI measurement is on.
Example
:CALC:TOI:STAT ON
CONFigure Commands
:CONFigure:MODE .............................................................. 95
:CONFigure:MODE
Set
Query
Description
Sets or queries the operating mode, spectrum or
power meter.
Query Syntax
:CONFigure:MODE {SA|PMETer}