Sense]:semask:carrier:cpsd, Sense]:semask:carrier:power – GW Instek GSP-930 Programming Manual User Manual
Page 146

GSP-930 Programming Manual
144
Syntax
[:SENSe]:SEMask:CARRier:AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:SEMask:CARRier:AUTO?
Parameter
0
Turn PSDRef/TotalPwrRef to manual
(off).
1
Turn PSDRef/TotalPwrRef to auto (on).
OFF
Turn PSDRef/TotalPwrRef to manual
(off).
ON
Turn PSDRef/TotalPwrRef to auto (on).
Return parameter 0
PSDRef/TotalPwrRef is set to manual.
1
PSDRef/TotalPwrRef is set to automatic.
Example
:SEM:CARR:AUTO OFF
[:SENSe]:SEMask:CARRier:CPSD
Set
Query
Description
Sets or queries the value of the PSDRef for SEM
measurements.
Syntax
[:SENSe]:SEMask:CARRier:CPSD <NR3>
Query Syntax
[:SENSe]:SEMask:CARRier:CPSD?
Parameter/
Return parameter
<NR3>
PSD ref unit = dBm/Hz
Example
:SEM:CARR:CPSD 20
[:SENSe]:SEMask:CARRier:POWer
Set
Query
Description
Sets or queries the value of the TotalPwrRef
amplitude for SEM measurements.
Syntax
[:SENSe]:SEMask:CARRier: POWer <ampl>
Query Syntax
[:SENSe]:SEMask:CARRier: POWer?
Parameter
<ampl>
<NRf>
Return parameter <NR3>
dBm
Example
:SEM:CARR:POW 2 dbm