Sense]:cnr:deltamarker:mode, Sense]:correction:cset<n>:data – GW Instek GSP-930 Programming Manual User Manual

Page 131

Advertising
background image

REMOTE CONTROL

129

Syntax

[:SENSe]:CNR:CHANnel:SPACe <freq>

Query Syntax

[:SENSe]:CNR:CHANnel:SPACe?

Parameter

<freq>

<NRf>

Return parameter <NR3>

Hz

Example

:CNR:CHAN:SPAC 6.0e+6

[:SENSe]:CNR:DELTamarker:MODE

Set

Description

Turns the CNR Noise Marking function to
Min(AUTO) or ΔMarker(MANual).

Syntax

[:SENSe]:CNR:DELTamarker:MODE {AUTO|MANual }

Parameter

AUTO

Sets the Noise Marking to Min.

ΔMarker

Sets the Noise Marking to ΔMarker.

Example

:CNR:DELT:MODE AUTO

[:SENSe]:CORRection:CSET<n>:DATA

Set

Query

Description

As a command, sets an offset for a certain
frequency for a selected correction set.
As a query, returns the data contents for the
selected correction set as <csv data>. The data will
be arranged as:
pt#1 freq, pt#1 offset, pt#2 freq, pt#2 offset,...

Syntax

[:SENSe]:CORRection:CSET<n>:DATA <freq>,<offset>

Query syntax

[:SENSe]:CORRection:CSET<n>:DATA?

Parameter

<freq>

<NRf> Hz

<offset>

<NRf> dB

<n>

<NR1>correction set number

Return parameter <CSV data> pt#1 freq, pt#1 offset,….. pt#n freq, pt#n

offset

Example

CORR:CSET1:DATA 2e+6,30

Advertising