Source:power[:level][:immediate] [:amplitude]:step, Source:power:mode – GW Instek GSP-930 Programming Manual User Manual
Page 165

REMOTE CONTROL
163
:SOURce:POWer[:LEVel][:IMMediate]
[:AMPLitude]:STEP
Set
Query
Description
Sets or queries the tracking generator step level.
Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP <rel_ampl>
Query Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP?
Parameter
<rel_ampl> <NRf>
Return parameter <NR3>
dB
Example
:SOUR:POW:STEP .5 db
:SOURce:POWer[:LEVel][:IMMediate]
[:AMPLitude]:STEP:AUTO
Set
Query
Description
Turns the tracking generator step level setting to
auto (on) or manual (off).
Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP:AUTO {OFF|ON|0|1}
Query Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP:AUTO?
Parameter
0
Turn TG step level to manual (off).
1
Turn TG step level to auto (on).
OFF
Turn TG step level to manual (off).
ON
Turn TG step level to auto (on).
Return parameter 0
TG step level is set to manual.
1
TG step level is set to automatic.
Example
:SOUR:POW:STEP:AUTO 1
:SOURce:POWer:MODE
Set
Query
Description
Sets the Power Sweep mode.