System:version:hardware, System:version:software – GW Instek GSP-930 Programming Manual User Manual
Page 173

REMOTE CONTROL
171
Warning
Do not perform this command if the update file is
not on the USB drive.
If the update file is not on the USB drive, it will
cause the instrument to continuously loop until a
USB drive with the appropriate update file is
inserted into the USB drive.
Syntax
:SYST:UPD
:SYSTem:VERSion:HARDware?
Query
Description
Returns the system firmware version.
Query Syntax
:SYSTem:VERSion:HARDware?
Return parameter <string>
T.1.X.X.X
Example
:SYST:VERS:HARD?
>T.1.0.0.0
:SYSTem:VERSion:SOFTware?
Query
Description
Returns the system software version.
Query Syntax
:SYSTem:VERSion:SOFTware?
Return parameter <string>
T1.00_2011.11.21_13
Where:
T1.00 = firmare version
2011 = year
11 = month
21 = day
13 = internal reference (not for
end-user)
Example
:SYST:VERS:SOFT?
> T1.00_2011.11.21_13