Calculate:marker<n>:fcount: resolution:auto, Calculate:marker<n>:fcount:state – GW Instek GSP-930 Programming Manual User Manual
Page 82

GSP-930 Programming Manual
80
Example
:CALC:MARK1:FCO:RES?
>1.0e+3
:CALCulate:MARKer<n>:FCOunt:
RESolution:AUTO
Set
Query
Description
Sets the frequency counter resolution Auto setting
on/off.
Syntax
:CALCulate:MARKer<n>:FCOunt:RESolution:AUTO
{ON|OFF|1|0}
Query Syntax
:CALCulate:MARKer<n>:FCOunt:RESolution:AUTO?
Parameter
<n>
<NR1>Marker number 1~6
0
Auto is off.
1
Auto is on.
OFF
Auto is off.
ON
Auto is on.
Return parameter 0
Auto is off.
1
Auto is on.
Example
:CALC:MARK1:FCO:RES:AUTO?
:CALCulate:MARKer<n>:FCOunt:STATe
Set
Query
Description
Sets or queries the state of the frequency counter
function.
Syntax
:CALCulate:MARKer<n>:FCOunt:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:MARKer<n>:FCOunt:STATe?
Parameter
<n>
<NR1>Marker number 1~6
0
Turn frequency counter off.
1
Turn frequency counter on.
OFF
Turn frequency counter off.
ON
Turn frequency counter on.
Return parameter 0
Frequency counter is off.
1
Frequency counter is on.
Example
:CALC:MARKer1:FCO:STAT 1