Calculate:demod:am:result:maximum, Calculate:demod:am:state – GW Instek GSP-930 Programming Manual User Manual
Page 74

GSP-930 Programming Manual
72
:CALCulate:DEMod:AM:RESult:MAXimum?
Query
Description
Returns the maximum recorded measurement
results for AM demodulation as a comma
separated string.
Query Syntax
:CALCulate:DEMod:AM:RESult:MAXimum?
Return parameter <depth,rate,power,offset,sinad>
depth
Modulation depth in %.
rate
Modulation rate in Hz.
power
Carrier power in the current Y-axis units.
offset
Carrier frequency offset in Hz.
sinad
Signal to noise and distortion ratio in dB.
Query Example
:CALC:DEM:AM:RES:MAX?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1
:CALCulate:DEMod:AM:STATe
Set
Query
Description
Sets or queries the state of the AM Analysis
function.
Syntax
:CALCulate:DEMod:AM:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:DEMod:AM:STATe?
Parameter
0
Turn AM Analysis off.
1
Turn AM Analysis on.
OFF
Turn AM Analysis off.
ON
Turn AM Analysis on.
Return parameter 0
AM Analysis is off.
1
AM Analysis is on.
Example
:CALC:DEM:AM:STAT 1