24. PIPETTE CALIBRATION

Balance allows pipette calibration by means of a special balance software or by

means of cooperation with computer software “Pipettes” intended for pipettes calibration

(workstation for pipettes calibration).

Before pipettes calibration a special set must be mounted inside the draft shield. The

set is an atypical balance equipment. See the picture below for set assembly instruction.

MYA 3Y balance with a set for pipettes calibration


Calibration set assembly procedure:

dismantle the weighing pan and the draft shield

inside the chamber place:

pan shield (1),

bottom ring (2),

balance pan (3),

glass vessel (5),

evaporation ring (6),

glass shield (7),

supplementary shield (8),

or glass shield (4),

XA 3Y balance with a set for pipettes calibration:


Calibration set assembly procedure:

dismantle the weighing pan and the draft shield,

place centering ring (8) inside the chamber,

put glass ring (7) onto the centering ring,

place weighing pan (6) inside the glass ring,

put top ring (5) onto the glass ring,

put evaporation ring (4) onto the top ring (5)

insert calibration vessel (3) into the evaporation ring,

put the glass cover (2) onto the top ring,

put glass supplementary cover (1) onto the glass cover.