7. Minimum volume: for specifying the minimum volume of a pipette (in

case of a pipette with fixed volume, insert <0>);

8. Number of channels: for specifying number of pipette’s channels, in

case of a single channel pipette insert <1>;

9. Type: NONE/A/D1/D2. Pipette type is compatible with applicable

standard. Selecting pipette’s type is mandatory if calibration process

complies with the requirements of standard ISO 8655, as values of

errors differ in each type of a pipette. Therefore, if the software is to

accept adequate error values it is necessary to determine type of a

pipette;

10. Tested volumes: if pressed, a window is opened with a list of

calibrated volumes (in case of a new pipette, the list is empty), add

volume to be tested, and define error value for each tested volume:

Press <

Add > soft key

A window with on-screen numeric keyboard is displayed,

Insert value of tested volume in [µl] and accept by pressing <

> key,

A new record is automatically added to the list with recommended

error values,

In order to change error values press a field with added volume

of a pipette,

each field is editable and enables inserting user values,

Caution:

While in calibration process, the order of tested volumes corresponds to

the order in which they are entered.

On setting correct values return to the main window.

24.4. Printouts

Option Printouts enables setting the content of a standard printout as well

as the non-standard printout.

Standard printout

comprises four internal sections featuring different variables. For each

variable set option YES to include it in a standard printout. If option NO is

selected, then the variable is disabled, i.e. it is not printed.