Acquiring part mass from database, Updating part mass in the database – RADWAG BALANCES 3Y SERIES User Manual

Page 74

Advertising
background image

- 74 -

Total mass of all parts loaded on the weighing pan must not be lower than value

declared in parameter “Minimal reference mass”. Unless this condition is met,

the balance displays a message: <Too low sample mass>;

Mass of a single part must not be lower than 0,1 of balance’s reading unit.

unless this condition is met, the balance displays a message: <Too low part

mass >.

16.5. Acquiring part mass from database

A product record in the database of products features set of information which

identifies it. One of them is mass, which is used during part counting process.

Procedure:

When in working mode < Parts counting> press <

Products database>

key, and then select desired product from displayed list.

Insert standard (reference) mass to balance’s memory

Procedure for adding standard (reference) mass of a single part to the

database of products:

a) Press < Setup > key followed by pressing < Database > key

b) In the Database of Products press < Products >

c) Press name of a product and edit data in field no. [5] Mass

d) Return to <Parts counting> mode.

If there are no data in the database:

a) Press < Setup > key followed by pressing < Database > key
b) In the Database of Products press <

Add > key

c) Accept the process of adding a new record in the database

d) Fill in field referring to a product, including field no. [5] Mass

e) Return to <Parts counting> mode.

16.6. Updating part mass in the database

Determined mass of a single part can be assigned to a product in the

database of products. This option is applied while using ACAI (Automatic

Accuracy Correction) function in order to determine part mass with high

accuracy.

Procedure:

a) Determine mass of a single part

b) Press <

Database> key

c) Press and hold finger on name of a product, which mass is updated,

d) Contextual menu is displayed,