Filter Very slow

Value release Reliable

Repeatability 0.0207 g

Stabilization time 5.015 s

Signature

........................................

PROCEDURE:

On function activation, the software initiates the process automatically. The

display indicates progress of the process course. On completing the process, the

software displays a summary and marked values of currently set filter values.

The report can be printed.

Caution:

Exiting the reporting window causes erasing the test results.

User can abort the process at optional moment by pressing <X> key in the

testing window.

AUTOTEST GLP

The test controls repeatability of placing the internal weight and

determining error of indication with reference to balance’s Max capacity.

Testing Procedure:

•

Two times loading of the internal weight, followed by 10 times

loading of the internal weight

•

Balance internal adjustment

•

Calculating and storing the value of standard deviation

•

In balances with automatically opened doors, door opening and

closing test is carried out.

Additionally the function enables:

•

Displaying and printing a report on available printers and saving

the report in balance’s memory. The report contains basic data on

balance’s parameters, ambient conditions and test results.

Test results:

*Deviation for Max.

*Value of repeatability of indication expressed as standard deviation

*Result of door test (positive / negative) – if a balance features

automatic door opening system.