Measuring units – RADWAG BALANCES 3Y SERIES User Manual

Page 63

Advertising
background image

- 63 -

Autozero function: Dosing

The function sets autozero operation mode to default one set for

dosing mode.

Accessible values:

NO - autozero operation is automatically disabled on entering

dosing mode

YES - autozero operation is automatically enabled on entering

dosing mode

Last digit

The function determines visibility of the last decimal place

indicated on measurement result. The functions provides three

settings:

- ALWAYS: al the digits visible

- NEVER: last digit invisible

- WHEN STABLE: last digit visible only for a stable

measurement

Ambient conditions

This parameter refers to ambient conditions of the balance. By default the

parameter is set to value Stable but it is advisable to set the

parameter to value Unstable for unstable ambient conditions.

Stable setting results in faster operation, i.e. weighing takes

much less time than in case of setting to value unstable.

Unstable

Stable

Measuring units

14.2.4.
Balance user can declare in a profile: start unit, supplementary unit and two

custom measuring units.

A custom measuring unit features:

o

A multiplier

o

A name (3 characters)

If a custom unit is designed, then its name is added to the list of accessible

measuring units.

This menu additionally enables inserting the value of gravitational acceleration

force for balance’s place of use. It is obligatory should a balance be used to

determine mass in [N].