Use the on-screen keyboard to insert desired value of a standard, or

Select desired value of a standard using a list displayed on pressing <

> key

Accept by pressing

key.

Caution:

Balance 3Y series cooperates with an additional display WD5/3Y manufactured by

RADWAG.

Ensuring correct cooperation between the balance and the additional display, requires
that the parameter <

Standard> has a variable number set to value {140}, and

the baud rate value set to 38400 bits/s for a port having the additional display

plugged.

31.6. Ambient conditions module

A balance 3Y series enables connecting an ambient conditions module THB series

using ports COM 1 or COM 2.

Ensuring correct cooperation between the balance and the ambient conditions

module requires entering module connection address and the baud rate of port

to which the THB module is plugged.

(address and baud rate settings are specified on module data plate).

31.7.

Tablet feeder (automatic feeder)

(

mode not available in balance’s standard version)

Balance 3Y series enables communicating it with an automatic feeder PA-02/H

series using ports:

RS 232 (COM1),

RS 232 (COM2),

Com internal.

Ensuring a correct cooperation with the feeder requires setting an appropriate

address of the device (the address is specified on the data plate of the automatic

feeder, including baud rate parameters).