RADWAG BALANCES 3Y SERIES User Manual

Page 213

Advertising
background image

- 213 -

Format: Z CR LF

Accessible responses:

Z_A CR LF

Z_D CR LF

- command understood and in progress

- command carried out

Z_A CR LF

Z_^ CR LF

- command understood and in progress

- command understood but zeroing range exceeded

Z_A CR LF

Z_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement

result

Z_I CR LF

- command understood but not accessible at this moment


Tare the balance

Format: T CR LF

Accessible responses:

T_A CR LF

T_D CR LF

- command understood and in progress

- command carried out

T_A CR LF

T_v CR LF

- command understood and in progress

- command understood but tarring range exceeded

T_A CR LF

T_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable

measurement result

T_I CR LF

- command understood but not accessible at this moment

Give tare value

Format: OT CR LF

Accessible responses: OT_TARA CR LF – command carried out

Response format:

1 2

3

4-12

13

14 15 16

17

18 19

O T

space

tare

space

unit

space

CR LF

Tare - 9 characters with right justification

Unit - 3 characters with left justification

Caution:

Tare value is always given in adjustment unit.

Set tare

Format: UT_TARE CR LF, where TARE – tare value

Accessible responses:

UT_OK CR LF - command carried out

UT_I CR LF

- command understood but not accessible at this moment

ES CR LF

- command not recognized (tare format incorrect)

Caution:

Use dot in tare format as decimal point.

Advertising