the weighing chamber and the measuring vessel, thus stopping liquid

evaporation process.

Carried out research proved that the adapter for pipettes calibration eliminates

or considerably reduces liquid evaporation process, which is essential for correct

gravimetric calibration of piston pipettes.

Selecting an appropriate adapter and balance mostly depends on type of

calibrated pipettes.

Before starting pipettes calibration process, assembly the adapter inside

balance’s weighing chamber, and insert data on pipettes to be calibrated,

including complete pipette characteristics and values of errors (see: Database of

Pipettes).

The following step is setting process’s main option:

Number of measurements

Request batch number (YES/NO)

Operation with ISO 8655 (YES/NO)

Acquire ambient conditions from THB module (YES/NO)

Caution:

Description of options and settings are given in point 24.1 of this user manual.