Print data from header

Print data from footer

Caution:

Measuring units for printing mass indication:

Net – main measuring unit (adjustment unit)

Tare – main measuring unit (adjustment unit)

Gross – main measuring unit (adjustment unit)

Current result – currently displayed measuring unit

Supplementary unit – supplementary measuring unit

Mass – main measuring unit (adjustment unit)

Non-standard printout

Non-standard printout

a printout may contain: TEXTS and VARIABLES(which are acquired from

the software on printout).

Each non-standard printout is a separate project, featuring specific name,

by which it is identified, and saved in the database of printouts.


PROCEDURE:

1. Press option <Non-standard printout>

2. Press <Add>

key

3. A window is opened with data such as Name/Code/Project
4. Set name and code of a printout
5. Press <Project> key

6. A keyboard is opened on the display for editing the printout

7. Use keyboard to design printout template, the printout may contain

texts and variables.

Caution:

User can add a printout by importing configured texts from data storage devices

connected to balance’s USB port.

Printout’s name is ONLY a NAME, and it is not part of its content.

Means of designing a non-standard printout template is described in point 24.11

<Printouts> of this user manual

15.6. Profiles

Profiles are described in point 14, Profiles, of this user manual.