Proximity sensors delay, Autotest, Proximity sensors delay 33.11. autotest – RADWAG BALANCES 3Y SERIES User Manual

Page 203

Advertising
background image

- 203 -

33.10. Proximity sensors delay

Parameter for setting delay of proximity sensors, it is expressed in [ms].

By default the value is set to <0>.

When using option of motion recognition the user must remember about

setting <proximity sensors delay> parameter to value <500ms>

Procedure:

a) Go to <

Other> submenu.

b) Select <Proximity sensors delay> parameterand wait for a dialog

window with an on-screen keyboard to open.

c) Enter value <500>.

d) Confirm. The program returns to menu window.

33.11. Autotest

AUTOTEST function is designed to aid a user in both assessing balance’s

operation and diagnosing the reasons for occurrence of errors in weighing

which exceed the maximum permissible values for a given balance

model.

By means of a simple, repeatable and fully documented way the function

enables optimizing balance’s settings to maintain the best possible

repeatability and weighing time at a workstation. The main purpose of

function operation is the possibility of monitoring the above mentioned

parameters at optional moment and saving records from carried out

tests.

The functions is divided into two modules:

AUTOTEST FILTER

AUTOTEST GLP

Before carrying out any of the tests , the balance controls its level status,

temperature and humidity.

AUTOTEST FILTER

It is process of 10 times loading and unloading of the internal weight for

all possible settings of balance’s filter and value release values. The test

controls 2 parameters:

Repeatability;

Stabilization time of measurement result.

The whole testing process takes approximately 1 hour. On completing the

test for all possible settings, balance displays a message box with

obtained results.

The user is given information on optimum balance settings in present

Advertising