MASS FORMAT - mass value in current measuring unit is returned

immediately

Frame format

1 2 3

4

5

6

7-15

16

17 18 19 20 21

S U I

stability

space sign mass space

unit

CR LF

Example:

S U I CR LF – command from a computer

S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out,

immediate response of mass value in current measuring unit

Where: _ - space

Switch on continuous transmission in basic measuring unit

Format: C1 CR LF

Accessible responses:

C1_I CR LF

- command understood but not accessible at this moment

C1_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in basic measuring unit

Frame format

1 2

3

4

5

6

7-15

16 17 18 19 20 21

S I space stability

marker space character mass space

unit

CR LF

Switch off continuous transmission in basic measuring unit

Format: C0 CR LF

Accessible responses:

C0_I CR LF

- command understood but not accessible at this moment

C0_A CR LF

- command understood and in progress

Switch on continuous transmission in current measuring unit

Format: CU1 CR LF

Accessible responses:

CU1_I CR LF

- command understood but not accessible at this moment

CU1_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in current measuring unit

Frame format:

1 2 3

4

5

6

7-15

16 17 18 19 20 21

S U I stability

marker space character mass space

unit

CR LF

Switch off continuous transmission in current weighing unit