37. COMMUNICATION PROTOCOL

General information

A. A character based communication protocol balance-terminal is designed for

establishing communication between a RADWAG balance and a peripheral

devices via RS 232 interface.

B. It consists of commands sent from a peripheral device to the balance and

responses from the balance..

C. Responses are sent from the balance on each receipt of a command as a

reaction for a specific command.

D. Commands forming the communication protocol enable obtaining data on

balance's status and influencing its operation, e.g.: request measurement

results from the balance, zero indication, etc.

37.1. A set of commands

Command Command description
Z

Zero balance

T

Tare balance

OT

Give tare value

UT

Set tare

S

Send stable measurement result in basic measuring unit

SI

Immediately send measurement result in basic measuring unit

SIA

Immediately send measurement results from both measuring

platforms basic weighing units

SU

Send stable measurement result in current measuring unit

SUI

Immediately send measurement result in current measuring unit

C1

Switch on continuous transmission in basic measuring unit

C0

Switch off continuous transmission in basic measuring unit

CU1

Switch on continuous transmission in current measuring unit

CU0

Switch off continuous transmission in current measuring unit

DH

Set low checkweighing threshold LO

UH

Set high checkweighing threshold HI

ODH

Give value of low checkweighing threshold LO

OUH

Give value of high checkweighing threshold HI

SM

Set mass value of a single item

TV

Set target mass value

RM

Set reference mass value

NB

Give balance factory number

PROFILE

Select balance profile

LOGIN

User log in

LOGOUT

User log out