RADWAG BALANCES 3Y SERIES User Manual

Page 114

Advertising
background image

- 114 -



Balance with a mounted pipette calibration

set.

Evaporation ring minimizes measurement errors being a result of liquid evaporation

occurring while weighing process.

Before pipettes calibration distilled water must be poured into evaporation ring, the

amount of water should not exceed 2/3 of the ring height.

The set may be operated after about an hour – this period of time is required for

moisture stabilisation. One must remember to control the level of distilled water –

surface of the vessel should remain covered with it. Excess of water may be removed by

means of automatic pump or external pipette.

In order to minimize both, changes of moisture content inside the weighing chamber

and negative effect of air blows during opening the chamber door, the liquid should be

dosed by means of pipette through an opening located in the top cover of the weighing

chamber.

It is allowed to use a balance without a draft shield for pipettes calibration. For this

reason both, the draft shield and frame must be dismounted (only XA.3Y.A series

balances). For dismount instruction of draft shield see chapter 2.6., for dismount

instruction of the frame see instruction below.