Minimum sample – RADWAG BALANCES 3Y SERIES User Manual

Page 47

Advertising
background image

- 47 -

Database of printouts enables selecting one of available non-

standard printouts visible in menu <PRINTOUTS>, or designing a

unique printout which is automatically added to the databases.

Caution: Means of designing printouts is described in point 15.5 of this

user manual.

Air buoyancy compensation

Contains parameters allowing the user to switch on the correction and

enter data relating to density of the sample and to density of the air.

CAUTION: function operates only for weighing mode.

For instruction and information on air buoyancy compensation operation

see point 10.7.

10.9. Minimum sample

The weighing mode settings comprise a function of <Minimum sample>.

The function is enabled on specifying in the database <Database/Minimum

sample > parameters on methods for determining the minimum sample

and the values of minimal sample weight specific for a method. In standard

version, this database item is not filled in with data.

Activities related to determining minimum sample and adding data

to the database: <Databases/Minimum samples> are accessible

only for an authorized “RADWAG” personnel.

Should balance operator use the minimum sample mode, and data on

minimal sample weight are not entered to balance menu, then please

contact the local representative of “RADWAG” company.

An authorized RADWAG representative determines the minimum sample

dependent on packaging mass. The process is carried out with mass

standards, in balance place of use and in accordance with the requirements

of the accepted quality system. The obtained values are entered to balance

software: <Databases/Minimum samples>.

A minimum sample determining method may have a few tare values

defined, along with assigned values of minimum samples, expiry dates of

the carried out measurements and entered data. The settings are not to be

edited by balance user.

Using <Minimum sample> mode guarantees that the results of completed

weighing processes are within set tolerance limits, and they comply with

the requirements of the quality system accepted by an organization.

CAUTION: the mode operates in the weighing mode only.