Pipettes calibration – quick access keys, Adding a pipette to the database of pipettes – RADWAG BALANCES 3Y SERIES User Manual

Page 118

Advertising
background image

- 118 -

<Comply with ISO 8655> the error values are neglected, and replaced by

those from ISO standard),

Acquire data on ambient conditions from THB module: YES/NO. For

option <YES> the software automatically, and in specified moment,

acquires and saves ambient conditions’ parameters from a connected THB

module. If option <NO> is selected, before starting and after completing

the pipettes calibration process operator should manually insert ambient

conditions parameters (i.e. temperature, humidity and pressure in the

weighing room) from external ambient conditions sensors.

Result control, Tare mode, Automatic footer printout, Printout

mode, Printout, means of using these functions is given in point 10.8.

“Additional parameters on weighing process”.

24.2. Pipettes calibration – quick access keys

Each of working modes features a set of default quick access keys which are

automatically displayed on mode activation. The set of keys can be modified

by selecting other quick access keys to the bottom bar of the display. Such

process requires appropriate operator’s access level.

24.3. Adding a pipette to the Database of Pipettes

Database of pipettes contains list of pipettes’ names and other data including

tested volume and errors values for a specific volume. When adding a pipette

to the database, first specify its name, and then add other data on a pipette.

The software operates intuitively, and commands a user to follow displayed

information.
Adding a pipette to the Database can be carried out in menu of the database.

Procedure:

Enter submenu < Databases>, press <

Pipettes> field

Press <

Add > soft key, should a new pipette be added

The software automatically adds a new entry to the database, and enters its

editing mode. Insert all parameters of a new formulation.

List of parameters defined for a pipette:

1. Name: press “Name” field, a window opens for inserting name of a

pipette;

2. Code: for inserting pipette’s code;

3. Model: for inserting model name of a pipette;

4. Tip: for inserting name of used tip;

5. Type of volume: ADJUSTABLE / FIXED;

6. Nominal volume: for specifying nominal volume of a pipette;