RADWAG BALANCES 3Y SERIES User Manual

Page 218

Advertising
background image

- 218 -

Accessible responses:

SM_OK CR LF

SM_I CR LF

- command carried out

- command understood but not accessible at this

moment (e.g.: other working mode, not PARTS

COUNTING)

ES CR LF

- command not recognized (incorrect mass format)

Set target mass value (e.g. for DOSING)

Format: TV_XXXXX CR LF, where: _ - space, XXXXX – mass format

Accessible responses:

TV_OK CR LF

TV_I CR LF

- command carried out

- command understood but not accessible at this

moment (e.g.: WEIGHING mode)

ES CR LF

- command not recognized (incorrect mass format)


Set reference mass value (e.g. for PERCENT SETUP)

Format: RM_XXXXX CR LF, where: _ - space, XXXXX – mass format

Accessible responses:

RM_OK CR LF

RM_I CR LF

- command carried out

- command understood but not accessible at this

moment (e.g.: other working mode, not PERCENT

SETUP)

ES CR LF

- command not recognized (incorrect mass format)


User log in

Format: LOGIN Name, Password CR LF (Name and password have to be

inserted as they are set in the balance – small and capital letters)

Accessible responses:

LOGIN OK CR LF

- command understood, new user is logged in

LOGIN ERRROR CR LF - command understood, an error in name or

password occurred, log in failed

ES CR LF

- command not recognized (error in format)

User log out

Format: LOGOUT CR LF

Accessible responses:

LOGOUT OK CR LF

- command understood, user is logged out

ES CR LF

- command not recognized (error in format)

Set balance mode

Format: PROFILE_Name CR LF where: _ - space (Name must be accordant

with the name form entered in a balance – lowercase and uppercase letters)

Accessible responses:

PROFILE OK CR LF

- command understood, new PROFILE activated

Advertising