RADWAG BALANCES 3Y SERIES User Manual

Page 179

Advertising
background image

- 179 -

2. Code – method code

3. Description – method description

4. Next control – a field for entering expiry date of the minimum sample.

2 weeks before the expiry date the pictogram on minimum sample

status displayed in the mass measuring window section is

supplemented by a clock. It informs on the approaching expiry of the

minimum sample. Contact RADWAG representative to carry out

necessary changes in minimum sample settings.

5. Limits – option that enables entering data on the mass value of a

minimum sample and range of packaging mass (tare) with applicable

criteria:

Tare – maximum value of tare for which the minimum sample is valid.

the software enables entering three characteristic quantities: 0.000g,

optional mass from balance measuring range and Max balance

measuring range (see a below example).

Minimum mass – the value of minimum mass that is determined for

a balance in its place of use and using appropriate methods.

Example no. 1 a balance AS 220.3Y, d=0.0001 g:

This balance has the following values of minimum sample determined for the

following values of packaging mass.

No. Tare value

Minimum

sample

Means of operation

1

0.0000 g

0.1000 g The minimum sample refers to all net

weighments that are weighed without a

packaging (the <TARE> button is not

applied)

2

10.0000 g

1.0000 g The minimum sample refers to all net

weighments that are weighed in a

packaging that mass ranges from 0.0001 g

to 10.0000 g inclusive (the <TARE> button

applied)

3

50.0000 g

2.5000 g The minimum sample refers to all net

weighments that are weighed in a

packaging that mass ranges from 10.0001

g to 50.0000 g inclusive (the <TARE>

button applied)

4

200.0000 g 4.0000 g The minimum sample refers to all net

weighments that are weighed in a

packaging that mass ranges from 50.0001

g to 200.0000 g inclusive (the <TARE>

button applied)