Application of outputs:

The set of outputs is used to SIGNAL the status of measurement result.

Change of logic status of the output system, e.g. from [0] to [1] takes place

on meeting a requirement assigned to the output.

Each of 4 outputs is connectable to the following data:

Available options

Balance status, which switches the logic of

output

None

-------

Stable

Each stable measurement switches the logic of

the system

MIN stable

Stable measurement below [MIN] threshold

switches the logic of the system,

MIN unstable

Unstable measurement below [MIN] threshold

switches the logic of the system,

OK stable

Stable measurement within [MIN] [MAX]

thresholds switches the logic of the system,

OK unstable

Unstable measurement within [MIN] [MAX]

thresholds switches the logic of the system,

MAX stable

Stable measurement below [MAX] threshold

switches the logic of the system,

MAX unstable

Unstable measurement below [MAX] threshold

switches the logic of the system

Confirmation of cycle

completion

End of procedure

Procedure


- Press

key and press: “ Inputs / Outputs” key.

- Enter submenu < Inputs / Outputs>, select and press <

Outputs> key

- Enter editing mode of a selected output, which opens list of available

functions,

- Select a function from displayed list and return to weighing with saving

carried out changes.