Parts counting procedure, Checkweighing function in parts counting mode – RADWAG BALANCES 3Y SERIES User Manual

Page 75

Advertising
background image

- 75 -

e) Select option <Assign standard>, reference mass is saved in a product

record under entry <Mass>.

16.7. Parts counting procedure

The first step in parts counting mode is obtaining data on mass of a single

part. Select one of available options:

Give mass value of a single part (see point 16.3) and place parts on

balance’s weighing pan, balance displays totalized parts.

Set mass of a single part from a given quantity of parts (see point

16.4.), the balance additionally shows pictogram of the ACAI function

<

> (if enabled). Place parts on balance’s weighing pan, the balance

displays totalized parts.

Acquire mass of a single part from database of products (see point

16.5.) by selecting a desired product record. Place parts on balance’s

weighing pan, the balance displays totalized parts.

Caution:

All additional elements (i.e. packaging) have to be tarred before starting the parts

counting process.

16.8. Checkweighing function in parts counting mode

Parts counting process can be aided by checkweighing function, i.e. control

whether indication is within set thresholds.

Checkweighing requires setting values of two thresholds, described as:

LOW threshold [Min= … parts]

HIGH threshold [Max= … parts]

and enabling bargraph (set its attribute to YES), which displays below

relation:

CURRRENT NO. OF PARTS / CHECKWEIGHING THRESHOLDS.

Defining the values of Low and High thresholds is carried out in the Database

of Products while editing Product entry, or using a quick access key <

Defining thresholds>.

Advertising