2.2. Balances XA 3Y; XA 3Y.A

A – balance with d=0,01mg unit
B – balance with d=0,1mg unit

After balance installation at operating

place, remove the transport protection (9) -

press the safety catch gently and turn it

accordingly to <OPEN> instruction, next

remove the safety catch but remember to

retain it should the device be transported in

the future.

Next, assembly the following parts:

bottom shield of the drying

chamber (5)

centering ring (4)

weighing pan (2)

pan shield (3) or (1)

After the elements setting, additional

devices should be connected and balance

should be plugged to power source. Power

socket is located on back side of balance

housing.

Pipettes calibration

Remove the pan and put on its place:

- pipettes calibration chamber

- inside the chamber, pan and other

elements should be set up.

Balances XA 3Y.F

After placing the balance in its intended

place of use, assembly all of its standard

components:

- bottom shield of the weighing

chamber (4)

- bottom ring (3)

- standard weighing pan (1)

- pan shield (2)

After assembling all of the components and

connecting peripheral devices plug the

balance to power source using a dedicated

power adapter. Power socket is located on

back side of balance housing.
Before weighing filters remove the pan

shield (2), the standard weighing pan (1)

and the bottom ring (3). Next, inside the

weighing chamber place a glass shield (6)

and assembly a pan-stand for weighing

filters (5).
on completing the above process zero/tare

balance indication. The system is ready

weighing filters procedures.

CAUTION! Any activities have to be carried out very carefully and gently

not to damage the measuring system of the balance.