Manual printout / automatic printout – RADWAG BALANCES 3Y SERIES User Manual

Page 223

Advertising
background image

- 223 -

n – parameter, decimal value specifying working mode number. See OMI

command description for detailed information.

Example

Command:

OMG<CR><LF>

– give current working mode

Response:

OMG_13_OK<CR><LF>

– the device currently operates in Statistics mode

Send all implemented commands

Format: PC CR LF

Accessible response:

PC_A_”Z,T,S,SI,SIA,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,

LOGIN,LOGOUT,PC" – command carried out, terminal has sent all

implemented commands.

37.3. Manual printout / Automatic printout

Balance 3Y series enables generating manual or automatic printouts.

•

Manual printout: press

key on stabilization of indication

(measurement result),

•

Automatic printout is generated automatically, in accordance to the settings

of automatic printout (see point 10.8 of this user manual).

The content of a printout depends on setting in menu <Standard printout> -

<Weighing printout template>(see point 15.5 of this user manual).

Format of mass printout:

1

2

3

4 -12

13

14 15 16 17 18

stability

marker

space character mass space

unit

CR LF

Stability marker

[space] if stable

[?] if unstable

[!] air buoyancy compensation mode is enabled

[^] if error of exceeding range to + occurs

[v] if error of exceeding range to – occurs

Character

[space] for positive values

[-] for negative values

Mass

9 characters with decimal point and right justification

Unit

3 characters with left justification

Example:

Advertising