n – parameter, decimal value specifying working mode number. See OMI

command description for detailed information.

Example

Command:

OMG<CR><LF>

– give current working mode

Response:

OMG_13_OK<CR><LF>

– the device currently operates in Statistics mode

Send all implemented commands

Format: PC CR LF

Accessible response:

PC_A_”Z,T,S,SI,SIA,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,

LOGIN,LOGOUT,PC" – command carried out, terminal has sent all

implemented commands.

37.3. Manual printout / Automatic printout

Balance 3Y series enables generating manual or automatic printouts.

Manual printout: press

key on stabilization of indication

(measurement result),

Automatic printout is generated automatically, in accordance to the settings

of automatic printout (see point 10.8 of this user manual).

The content of a printout depends on setting in menu <Standard printout> -

<Weighing printout template>(see point 15.5 of this user manual).

Format of mass printout:

1

2

3

4 -12

13

14 15 16 17 18

stability

marker

space character mass space

unit

CR LF

Stability marker

[space] if stable

[?] if unstable

[!] air buoyancy compensation mode is enabled

[^] if error of exceeding range to + occurs

[v] if error of exceeding range to – occurs

Character

[space] for positive values

[-] for negative values

Mass

9 characters with decimal point and right justification

Unit

3 characters with left justification

Example: