RADWAG BALANCES 3Y SERIES User Manual

Page 62

Advertising
background image

- 62 -

Balance user can influence to some extend the range of signal processing

by selecting an appropriate FILTER value. Available options:

o

very fast

o

fast

o

average / normal

o

slow

o

very slow

While setting the filtering level the user should consider the actual

operating conditions of a balance. In case of very good operating

conditions set the filter to average/normal or fast, in case of rough

conditions set the filter to slow or very slow.

Caution:

in case of precision balances 3Y series the recommended range of filter

is between very fast

÷

average/normal;

in case of analytical balances and microbalances 3Y series the

recommended range of filter is between average/normal

÷

very slow.

Value release

This parameter determines displaying the stability marker for a

measurement result.

There are 3 available settings of value release parameter

o

fast

o

fast + reliable /recommended/

o

reliable

Caution:

The speed of stability marker occurrence depends on applied filter value and

applied value release

Autozero function

The function is to automatically monitor and correct zero

indication of a balance.

If the function is enabled, the following measurement results are

compared to each other in constant time intervals. If the results differ

less than declared AUTOZERO range, e.g. 1 division, the balance is

automatically zeroed, and the markers of stable indication

and

precise zero

are displayed.

Active AUTOZERO means, that each measurement starts from the precise

zero point. There are, however, cases where the function may disturb the

weighing process. For instance during very slow load placing or pouring

onto the balance’s weighing pan. In such case, the correcting system of

zero indication may also correct the actual indication of a load placed on

the weighing pan.

Accessible settings: NO - autozero function disabled

YES - autozero function enabled

Advertising