RADWAG BALANCES 3Y SERIES User Manual

Page 221

Advertising
background image

- 221 -

Command:

BP_350<CR><LF> – turn BEEP signal for 350 ms

Response:

BP_OK<CR><LF> – BEEP on

Caution! – BEEP signal activated by BP command gets aborted if it is activated

by means of other devices: keyboard, touch panel, proximity sensors.

OMI - list available working modes

The command lists available working modes for a given device

Format: OMI <CR><LF>

Accessible responses:

OMI <CR><LF>

n_”Mode name”

<CR><LF>

:

n_”Mode name”

<CR><LF>

OK <CR><LF>

– command completed, available working modes are

given

OMI_I <CR><LF>

– command understood but not accessible at this

moment

Mode name – parameter, working mode name, given in between inverted

commas. Form of the name is the same like on the balance display, it is given

in a currently operating language.

n – parameter, decimal value specifying working mode number.

n → 1 – Weighing

2 – Parts counting

3 – Percent setup

4 – Dosing

5 – Formulation

6 – Animal weighing

7 – Density

8 – Density of solids

9 – Density of liquids

10 – Peak hold

11 – Totalizing

12 – Checkweighing

13 – Statistics

14 – Pippette calibration

Caution: working modes numbering is assigned to working modes names and

it is the same for each balance types.

Some balances in response to OMI may give working modes numbers without

names.

Example1

Command:

OMI <CR><LF>

– give available working modes