Activating procedure – RADWAG BALANCES 3Y SERIES User Manual

Page 120

Advertising
background image

- 120 -

Settings for header printout, weighing printout, and footer printout are

specified in point. 15.5; below description refers to setting a printout

template for report on pipette calibration.

Content of each of reports:

Working mode

User

Client

Pipette

Serial no.

No. of channels

Channel no.

Number of measurements

Operation with ISO 8655

Start date

End date

Water temperature

Temperature

Humidity

Pressure

Z coefficient

Measurements and statistics

Statistics

Status

Empty line

Dashes

Signature

Non-standard printout

24.5. Activating procedure

Carrying out pipettes calibration process requires installing an adapter for

pipettes calibration. The adapter does not come standard with the balance.

The adapter for pipettes calibration is a simple supplementary device dedicated

for calibration and checking of piston pipettes. It is designed for application on

balances MYA 21.3Y and XA 52.3Y.

The adapter has been designed to minimize evaporation occurring in liquid

handling procedures. It consists of a minimized weighing chamber installed

inside the weighing chamber of a balance. The adapter features an evaporation

ring and a dedicated weighing pan for central locating of the measuring vessel.

Application of the adapter for pipettes calibration prevents from liquid

evaporation during calibration process. Small dimensions of the adapter,

combined with the evaporation ring enable maintaining high humidity rate inside

Advertising