10.3. Balance leveling

Balance Y2 series features AutoLEVEL System, which continuously monitors balance’s level

status during its operation. The level status is signaled in the upper right corner of balance’s

display. The system continuously controls level, and on detecting level change it indicates

appropriate information on the display and/or signals an alarm. Then the software opens a

window for adjusting the level.

Balance leveling procedure

Press level status pictogram<

> located in the upper right corner of the

display.

The display indicates a control panel of leveling function. Balance operator

previews level status, and balance’s top elevation.

Level the balance by turning the adjustable feet in a way indicated by

blinking pictograms on the display < > – the level point should move

towards the center of the level circle.

As the level point is in the center point, its colour changes from red to

green – which is a confirmation for correct level status.

CAUTION:

The balance is equipped with Automatic Level Control mechanism. Description to

be found in point 33.7 of this manual.

10.4. Zeroing

In order to zero mass indication press

key.

The mass indication on the display should equal zero, and precise zero

and

stability

markers should appear.

Zeroing process is tantamount to determining new zero point recognized by

the balance as precise zero. Zeroing is possible only in stable status of display

indication.

Caution:

Zeroing the display indication is possible only within the ±2% range of

instrument’s maximal capacity. If zeroed value is above the ±2% of the

maximal capacity, then the software indicates an error message Err2.

Advertising