Proximity sensors

15.3.1.

The balance comprises two proximity sensors enabling wave control of the balance therefore

the user does not have to either press any key on the overlay or touch the screen.

The program recognizes 4 types of motions performed near the sensor:

1. Placing hand near the left sensor <Left proximity sensor>

2. Placing hand near the right sensor <Right proximity sensor>

3. Motion to the left <Proximity sensor: left motion>

4. Motion to the right <Proximity sensor: right motion>

CAUTION: using option LEFT MOTION or RIGHT MOTION, it is necessary to

remember that <Proximity sensors delay> parameter must be set to value 500

ms!

To each of the motions, one of the available options may be assigned: [None; Profile;

Adjustment; Zeroing; Tarring; Tare setting; Tare switch off; Reset tare; Packaging; Print;

Header printing; Footer printing; Unit; Universal Variable 1; Universal Variable 2; Universal

Variable 3; Universal Variable 4; Universal Variable 5; Confirm; Abort; User; Left door; Right

door; Open/Close door; Parameters; Product, Warehouse; Client].

The selected option is recognized by the program and performed when activated by a

particular motion.

In order to ensure correct operation the user must remember to set the right sensors

sensitivity and the right value for proximity sensors delay (for more information refer

to point 33of this manual).

Automatically opened door

15.3.2.

This menu section comprises settings on automatic opening and closing

door of a balance featuring such mechanism, i.e. balances MYA 3Y and XA

3Y.A series. The settings may refer to the quick access keys as well as the

infrared proximity sensors located on the overlay of balance’s terminal.

Settings for a balance MYA 3Y series cover:

o

Left door / the option enables changing the position of the door on the left

side of the weighing chamber

o

Right door / the option enables changing the position of the door on the

left side of the weighing chamber

o

Open / Close door / the option enables changing the position of the door,

e.g. if the weighing chamber is opened, then activating the option causes

closing it. If the weighing chamber is closed, then activating the option

causes opening the door located on the right side of the weighing chamber,

Settings for a balance XA 3Y.A series cover:

o

Open / the option enables opening the door of the weighing chamber

o

Close / the option enables closing the door of the weighing chamber

o

Open / Close door / the option enables changing the position of the door,

e.g. if the weighing chamber is opened, then activating the option causes

closing it. If the weighing chamber is closed, then activating the option

causes opening the door of the weighing chamber.